209 Specifications

0-10 VDC Output Signal
Load must be at least 5 Kohms
0-5 VDC Output Signal
Load must be at least 2.5 Kohms
4-20 mA Output Signal
Load must not exceed 500 ohms
Accuracy (including linearity)
±1.5% Full Scale
Certifications
CE Approved
89/336/EEC (EN 55011 & EN 50082-1)
73/32/EEC Low Voltage Directive
Compatible Gases
Clean, dry gases compatible with wetted materials
Electrical Connections
Integrated 7-pin power/signal connector
Fitting Material
Choose from acetal, brass, or stainless steel
Leak Integrity
1 x 10-4 sccs of He
O-Ring Material
Viton
Power
12 VDC Models: 12 VDC @ 150 mA (12.5-15 VDC)
24 VDC Models: 24 VDC @ 80 mA (22-25 VDC)
Pressure Drop
No more than 0.075 psid (5 mbar)
Pressure Rating
150 psig (10.3 bar)
Pressure Sensitivity
±0.02% F.S. or less per psi (per 69 mbar)
Ratings
IP67 (NEMA 4)
Recommended Filtration
100 microns or less
Reliability
100,000 Hours MTBF (testing ongoing)
Repeatability
±0.50% Full Scale
Settling Time
Typically <1 second for 97% of final value
Temperature Rating
Operating Range: 0 to 50șC
Storage Range: 0 to 70șC
Temperature Sensitivity
±0.15% F.S. or less per șC
Warm-Up Time
Less than 5 minutes
Wetted Materials
Aluminum
304 Stainless Steel
316 Stainless Steel



ZERO STABILITY



FITTING CHART