Model 104
Quick Ship Units come standard with:
- Stainless Steel Bearing Supports
- 12 VDC power configuration
- Viton O-ring seals
- Calibrated for Deionized Water
- Acetal Tube Fittings
- Stainless Steel Tube Fittings
McMillan Model 104 FLO-SENSORS are capable of measuring extremely low liquid flow rates from 13 mL/minute up to 10 L/minute with a full scale accuracy of ±1.0% or better!
A wide variety of fluids may be measured. Repeatable results are achieved using a patented Pelton-type microturbine wheel. This proven design has been providing precision results since 1988 and has developed a well-deserved reputation for continuous operational service for many years without failure.
Because of the compact size and economical cost of these products, the Model 104 FLO-SENSORS are suitable for a wide variety of industrial, commercial, laboratory and O.E.M. applications. Some sample applications include measurement of hydrocarbon fluids, fuels, light oils, solvents, coolant, pesticides, mild acids, alkalis, and deionized water.
Several power and output configurations are available, including both pulse and analog outputs. NIST Traceable certificates are available.
A power adapter or cable assembly is required for operation.
104 Features & Options
ACCURACY/LINEARITY
All models have a standard accuracy specification of ±1.0% F.S. (including linearity). An improved accuracy specification of ±0.5% F.S. is available. NIST traceable calibration certifications are standard for improved accuracy (suffix H) models and optional for standard units.
DISPLAYS
McMillan has a comprehensive range of FLO-METERS with integrated displays. A number of remote displays are also available for use with any FLO-SENSOR.
ELECTRICAL CONNECTIONS
Most units have an integrated 4-pin male connector. To complete connections, either a cable assembly or power adapter should be ordered. Units where the circuit board has been epoxy potted for increased chemical resistance feature an integrated cable with pigtail leads.
FLOW RANGES
Flow ranges from 13-100 mL/minute up to 1.0-10.0 L/minute are available. Consult the factory for custom requirements.
FLUID CONNECTIONS
All units have compression type tube fittings as standard. Many alternate fitting types and sizes may be selected as allowed in the Product Configurator.
POWER
Most units may be specified to operate with either 12VDC or 24 VDC power. Various power adapters are also available for use with 12VDC versions.
SIGNAL OUTPUTS
Most units may be ordered with a 0-5VDC output or with both 0-5VDC and pulse outputs.
Pressure Drop
Fitting Codes
Specifications
- Accuracy (including linearity, best fit straight line):
- ±1.0% Full Scale
- Repeatability:
- ±0.2% Full Scale
- Pressure Rating:
- 500 psig (34 bar)
- Temperature Rating (Environment):
- Operating Range: 5 to 55ºC
Storage Range: 0 to 70ºC - Temperature Sensitivity:
- ±0.2% F.S. per ºC
- Wetted Materials:
- Stainless Steel
Ryton
Epoxy
Glass
Sapphire - O-Ring Material:
- Viton
- Fitting Material:
- Stainless Steel
- Recommended Filtration:
- 25 microns or less
- Compatible Liquids:
- Low viscosity (<10 cS)
Translucent or transparent
Minimum amount of entrained air - Output Signal:
- Non-isolated
2500 ohm minimum load - Power Requirements:
- 12VDC @ 35 mA (11.5-15 VDC)
- Typical Response Time:
- Typically <1 second for 63% of final value
- Reliability:
- 100,000 Hours MTBF
- Approvals/Certifications:
- CE Approved
89/336/EEC (EN 55011 & EN 50082-1)
73/23/EEC Low Voltage Directive - Ratings:
- IP10 (NEMA 1)
- Warranty:
- Model 104: 1 Year Limited
|
||||
|
||||
|
||||
|
||||
|
||||
|
||||
|
||||
|
||||
|
||||
|
||||
|
||||
|
||||
|
||||
|